Home / Products / Toga GC 2
Toga GC 2
Energy Support

Toga GC 2

High-sensitivity laboratory solution with vacuum degassing – ideal for acceptance, service and heat-run tests.


Product Overview

The fully automatic gas-in-oil analysis Toga GC 2 uses mercury-free vacuum degassing and analyses all 11 dissolved gases as well as the total gas content to ASTM D3612 and IEC 60567. This lets you detect gas formation from thermal faults or electrical discharges early and ensure your critical infrastructure is ready for the demands of today and tomorrow.

Key Specifications

Vacuum degassing : Multi-stage diaphragm pump without mercury
Carry-over : None – fully automatic cleaning after degassing
Vacuum sensor : 0.1 – 200 mbar
Sample volume : 30 – 50 ml
Gas chromatograph : TCD + FID + methanizer, 2 micro-packed columns
Autosampler (option) : For 16 or 32 syringes (30 / 50 ml)
Calibration : One certified gas standard for new and used oils
Power supply : 100 – 240 V AC, 50/60 Hz
Power consumption : < 500 W
Weight : 28 kg
Dimensions : 520 × 330 × 370 mm